Edge Ultrasound System Guida Utente Pagina 173

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
Vedere la pagina 172
Chapter 8: Acoustic Output 161
Acoustic Output
(a)This index is not required for this operating mode; value is <1.
(b)This transducer is not intended for transcranial or neonatal cephalic uses.
# No data are reported for this operating condition since the global maximum index value is not reported for the reason
listed. (Reference Global Maximum Index Value line.)
Data are not applicable for this transducer/mode.
Table 38: Transducer Model:
TEEx
Operating Mode:
PW Doppler
Index Label M.I.
TIS TIB
TIC
Scan
Non-scan
Non-scan
A
aprt
1A
aprt
>1
Global Maximum Index Value (a) (a) 1.7 (b)
Associated Acoustic
Parameter
p
r.3
(MPa) #
W
0
(mW) —# 29.29 #
min of
[W
.3
(z
1
),I
TA.3
(z
1
)]
(mW)
z
1
(cm)
z
bp
(cm)
z
sp
(cm) # 0.6
d
eq
(z
sp
) (cm) 0.34
f
c
(MHz) # # 3.84 #
Dim of A
aprt
X(cm) —# 0.261#
Y(cm)
—# 0.9 #
Other Information
PD (μsec) #
PRF (Hz) #
p
r
@PII
max
(MPa) #
d
eq
@Pll
max
(cm) 0.34
Focal Length FL
x
(cm) —# #
FL
y
(cm) —# #
I
PA.3
@MI
max
(W/cm
2
)
#
Operating
Control
Conditions
Control 1: Exam Type Crd
Control 2: Sample Volume 1mm
Control 3: PRF 2604
Control 4: Sample Volume Position Zone 1
Vedere la pagina 172
1 2 ... 168 169 170 171 172 173 174 175 176 177 178 ... 197 198

Commenti su questo manuale

Nessun commento