Chapter 8: Acoustic Output 131
Acoustic Output
(a)This index is not required for this operating mode; value is <1.
(b)This transducer is not intended for transcranial or neonatal cephalic uses.
# No data are reported for this operating condition since the global maximum index value is not reported for the reason
listed. (Reference Global Maximum Index Value line.)
Data are not applicable for this transducer/mode.
Table 8: Transducer Model:
C60x
Operating Mode:
2D
Index Label M.I.
TIS TIB
TIC
Scan
Non-scan
Non-scan
A
aprt
1A
aprt
>1
Global Maximum Index Value 1.0 (a) (b)
Associated Acoustic
Parameter
p
r.3
(MPa) 1.69
W
0
(mW) #— —#
min of
[W
.3
(z
1
),I
TA.3
(z
1
)]
(mW)
z
1
(cm)
z
bp
(cm)
z
sp
(cm) 4.7
d
eq
(z
sp
) (cm)
f
c
(MHz) 2.84 # #
Dim of A
aprt
X(cm) #— #
Y(cm)
#— #
Other Information
PD (μsec) 0.579
PRF (Hz) 5440
p
r
@PII
max
(MPa) 2.679
d
eq
@Pll
max
(cm)
Focal Length FL
x
(cm) #— #
FL
y
(cm) #— #
I
PA.3
@MI
max
(W/cm
2
)
197.7
Operating
Control
Conditions
Control 1: Exam Type Abd
Control 2: Optimization Res/
Gen
Control 3: Depth 11/
13 cm
Control 4: THI On
Control 5: MB (Multi Beam) On